Neonatal Pulse Oximetry Sensor Study
Accuracy and Precision of Peripheral Capillary Oxygen Saturation of Reprocessed Pulse Oximetry Sensors Compared to Oxygen Saturation in Arterial Blood Samples Assessed by CO-oximetry in Neonates
1 other identifier
interventional
23
1 country
2
Brief Summary
To validate the SpO2 accuracy, bias, and precision of Medline's reprocessed pulse oximetry sensors as compared to SaO2 in arterial blood samples as assessed by CO-oximetry in neonates.
Trial Health
Trial Health Score
Automated assessment based on enrollment pace, timeline, and geographic reach
participants targeted
Target at below P25 for not_applicable
Started Jun 2022
2 active sites
Health score is calculated from publicly available data and should be used for screening purposes only.
Trial Relationships
Click on a node to explore related trials.
Study Timeline
Key milestones and dates
First Submitted
Initial submission to the registry
February 10, 2022
CompletedFirst Posted
Study publicly available on registry
February 22, 2022
CompletedStudy Start
First participant enrolled
June 13, 2022
CompletedPrimary Completion
Last participant's last visit for primary outcome
October 10, 2023
CompletedStudy Completion
Last participant's last visit for all outcomes
October 10, 2023
CompletedResults Posted
Study results publicly available
October 30, 2024
CompletedOctober 30, 2024
October 1, 2024
1.3 years
February 10, 2022
October 2, 2024
October 29, 2024
Conditions
Outcome Measures
Primary Outcomes (1)
Bland-Altman Analysis for Nellcor Pulse Oximeter
BIAS Measurement and 95% confidence interval calculated using a Bland-Altman plot (mean vs. difference of SaO2 and SpO2) and an error plot (SaO2 vs. difference of SaO2 and SpO2) were generated. The B\&A plot analysis is a simple way to evaluate a bias between the mean differences, and to estimate an agreement interval, within which 95% of the differences of the second method, compared to the first one, fall. Data can be analyzed both as unit differences plot and as percentage differences plot. We ran a percentage mean difference plot.
15 episodic minutes
Study Arms (1)
Pulse oximeter sensor readings
OTHERSubjects who satisfy all of the inclusion criteria will be eligible to participate in the study, will receive intervention (the Masimo and Nellcor sensors, and their arterial blood sample will be taken in the normal course of care). This is a single arm study. Those who don't meet the inclusion/ exclusion criteria won't be eligible to participate.
Interventions
Pulse oximeter reading will be compared to CO-oximeter reading.
Eligibility Criteria
You may qualify if:
- Subjects who are postnatal 28 days or younger (neonates)
- Subjects who are likely to receive one or more ABG measurements as part of their clinical SOC
- Subjects who weigh less than 5 Kg (weight range for thesensors)
You may not qualify if:
- Subjects with current signs and symptoms of a clinically significant Patent Ductus Arteriosus (PDA) combined with a current or planned arterial line placement which will affect the validity of the co-oximetry measurement
- Subjects with physical malformation of hands, fingers, feet, or toes that would limit the ability to place sensors for this study
- Subjects judged by the Principal Investigator (PI) to be inappropriate for participation in this study
- Subjects for whom placing a pulse oximeter will cause dermatological issues (e.g. allergic to foam rubber or adhesive tape)
Contact the study team to confirm eligibility.
Sponsors & Collaborators
Study Sites (2)
Children's Hospital Los Angeles
Los Angeles, California, 90027, United States
Eastern Carolina University
Greenville, North Carolina, 27858, United States
Related Publications (3)
Ascha M, Bhattacharyya A, Ramos JA, Tonelli AR. Pulse Oximetry and Arterial Oxygen Saturation during Cardiopulmonary Exercise Testing. Med Sci Sports Exerc. 2018 Oct;50(10):1992-1997. doi: 10.1249/MSS.0000000000001658.
PMID: 29771822BACKGROUNDDawson JA, Davis PG, O'Donnell CP, Kamlin CO, Morley CJ. Pulse oximetry for monitoring infants in the delivery room: a review. Arch Dis Child Fetal Neonatal Ed. 2007 Jan;92(1):F4-7. doi: 10.1136/adc.2006.102749. No abstract available.
PMID: 17185428BACKGROUNDBland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571-82. doi: 10.1080/10543400701329422.
PMID: 17613642BACKGROUND
MeSH Terms
Conditions
Condition Hierarchy (Ancestors)
Results Point of Contact
- Title
- Julie J. Finley
- Organization
- Medline Industries, LP
Publication Agreements
- PI is Sponsor Employee
- No
- Restriction Type
- LTE60
- Restrictive Agreement
- Yes
Study Design
- Study Type
- interventional
- Phase
- not applicable
- Allocation
- NA
- Masking
- NONE
- Purpose
- DIAGNOSTIC
- Intervention Model
- SINGLE GROUP
- Sponsor Type
- INDUSTRY
- Responsible Party
- SPONSOR
Study Record Dates
First Submitted
February 10, 2022
First Posted
February 22, 2022
Study Start
June 13, 2022
Primary Completion
October 10, 2023
Study Completion
October 10, 2023
Last Updated
October 30, 2024
Results First Posted
October 30, 2024
Record last verified: 2024-10
Data Sharing
- IPD Sharing
- Will not share