Clinical Evaluation of CARESTREAM Dual Energy and Digital Tomosynthesis
1 other identifier
interventional
66
0 countries
N/A
Brief Summary
The purpose of this study is to evaluate the imaging performance of Dual Energy and Digital Tomosynthesis (DE-DT) application against the commercially available predicate devices.
Trial Health
Trial Health Score
Automated assessment based on enrollment pace, timeline, and geographic reach
participants targeted
Target at P50-P75 for not_applicable
Started Nov 2015
Shorter than P25 for not_applicable
Health score is calculated from publicly available data and should be used for screening purposes only.
Trial Relationships
Click on a node to explore related trials.
Study Timeline
Key milestones and dates
First Submitted
Initial submission to the registry
August 20, 2015
CompletedFirst Posted
Study publicly available on registry
August 24, 2015
CompletedStudy Start
First participant enrolled
November 1, 2015
CompletedPrimary Completion
Last participant's last visit for primary outcome
April 1, 2016
CompletedStudy Completion
Last participant's last visit for all outcomes
April 1, 2016
CompletedResults Posted
Study results publicly available
November 2, 2016
CompletedMarch 9, 2017
August 1, 2015
5 months
August 20, 2015
September 14, 2016
January 27, 2017
Conditions
Keywords
Outcome Measures
Primary Outcomes (9)
Radlex Scale for Diagnostic Quality Ratings - DE Predicate PA Chest
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
3 months
Radlex Scale for Diagnostic Quality Ratings - DE Investigational Composite
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
3 months
Radlex Scale for Diagnostic Quality Ratings - DE Investigational High Energy
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
3 months
Radlex Scale for Diagnostic Quality Ratings - DE Investigational Low Energy
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
3 months
Radlex Scale for Diagnostic Quality Ratings - DT Reference 1- PA Chest
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
3 months
Radlex Scale for Diagnostic Quality Ratings - DT Reference 2 - PA and LAT Chest
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
3 months
Radlex Scale for Diagnostic Quality Ratings - DT Investigational - Scout & DT Volume
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
3 month
Radlex Scale for Diagnostic Quality Ratings - LT Predicate Phantom Images
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
3 months
Radlex Scale for Diagnostic Quality Ratings - DT Investigational Phantom Images
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
3 months
Study Arms (3)
Predicate & Invest. DE - Human Subjects
EXPERIMENTALRadiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue).
Predicate & Invest. DT - Human Subjects
EXPERIMENTALRadiation - Fifteen to twenty (15-20) patients will receive a DR standard of care chest exam using the DRX Plus detector, and a DT exam. Each DT patient exam includes a scout image (chest PA) and a DT scan using the investigational DT SW. The DT scan is used by the DT console software to generate tomographic images.
Predicate & Invest. DT - Phantom Images
EXPERIMENTALRadiation - Eleven (11) phantoms of various anatomy will be imaged with linear tomography (LT) as predicate and DT for investigational.
Interventions
Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure. Phantoms were imaged using linear tomography and digital tomography.
Eligibility Criteria
You may qualify if:
- Age 18 years and older. Subject has provided informed consent. Study participant is scheduled to have a chest CT exam. Study participant is able to stand and be still during the exams.
You may not qualify if:
- Human subject specific:
- Not able or willing to provide Informed Consent, or consent is withdrawn. Study participant is a female of child bearing age.
Contact the study team to confirm eligibility.
Sponsors & Collaborators
MeSH Terms
Conditions
Interventions
Condition Hierarchy (Ancestors)
Intervention Hierarchy (Ancestors)
Results Point of Contact
- Title
- Susan Pate, RN, MS, RAC, Senior Clinical Manager
- Organization
- Carestream Health, Inc.
Study Officials
- PRINCIPAL INVESTIGATOR
Narinder Paul, MD
Toronto General Hospital
Publication Agreements
- PI is Sponsor Employee
- No
- Restrictive Agreement
- No
Study Design
- Study Type
- interventional
- Phase
- not applicable
- Allocation
- RANDOMIZED
- Masking
- DOUBLE
- Who Masked
- PARTICIPANT, OUTCOMES ASSESSOR
- Purpose
- DIAGNOSTIC
- Intervention Model
- PARALLEL
- Sponsor Type
- INDUSTRY
- Responsible Party
- SPONSOR
Study Record Dates
First Submitted
August 20, 2015
First Posted
August 24, 2015
Study Start
November 1, 2015
Primary Completion
April 1, 2016
Study Completion
April 1, 2016
Last Updated
March 9, 2017
Results First Posted
November 2, 2016
Record last verified: 2015-08
Data Sharing
- IPD Sharing
- Will not share