NCT02531646

Brief Summary

The purpose of this study is to evaluate the imaging performance of Dual Energy and Digital Tomosynthesis (DE-DT) application against the commercially available predicate devices.

Trial Health

100
On Track

Trial Health Score

Automated assessment based on enrollment pace, timeline, and geographic reach

Enrollment
66

participants targeted

Target at P50-P75 for not_applicable

Timeline
Completed

Started Nov 2015

Shorter than P25 for not_applicable

Status
completed

Health score is calculated from publicly available data and should be used for screening purposes only.

Trial Relationships

Click on a node to explore related trials.

Study Timeline

Key milestones and dates

First Submitted

Initial submission to the registry

August 20, 2015

Completed
4 days until next milestone

First Posted

Study publicly available on registry

August 24, 2015

Completed
2 months until next milestone

Study Start

First participant enrolled

November 1, 2015

Completed
5 months until next milestone

Primary Completion

Last participant's last visit for primary outcome

April 1, 2016

Completed
Same day until next milestone

Study Completion

Last participant's last visit for all outcomes

April 1, 2016

Completed
7 months until next milestone

Results Posted

Study results publicly available

November 2, 2016

Completed
Last Updated

March 9, 2017

Status Verified

August 1, 2015

Enrollment Period

5 months

First QC Date

August 20, 2015

Results QC Date

September 14, 2016

Last Update Submit

January 27, 2017

Conditions

Keywords

pulmonary nodule

Outcome Measures

Primary Outcomes (9)

  • Radlex Scale for Diagnostic Quality Ratings - DE Predicate PA Chest

    1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

    3 months

  • Radlex Scale for Diagnostic Quality Ratings - DE Investigational Composite

    1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

    3 months

  • Radlex Scale for Diagnostic Quality Ratings - DE Investigational High Energy

    1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

    3 months

  • Radlex Scale for Diagnostic Quality Ratings - DE Investigational Low Energy

    1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

    3 months

  • Radlex Scale for Diagnostic Quality Ratings - DT Reference 1- PA Chest

    1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

    3 months

  • Radlex Scale for Diagnostic Quality Ratings - DT Reference 2 - PA and LAT Chest

    1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

    3 months

  • Radlex Scale for Diagnostic Quality Ratings - DT Investigational - Scout & DT Volume

    1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

    3 month

  • Radlex Scale for Diagnostic Quality Ratings - LT Predicate Phantom Images

    1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

    3 months

  • Radlex Scale for Diagnostic Quality Ratings - DT Investigational Phantom Images

    1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.

    3 months

Study Arms (3)

Predicate & Invest. DE - Human Subjects

EXPERIMENTAL

Radiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue).

Radiation: Radiation

Predicate & Invest. DT - Human Subjects

EXPERIMENTAL

Radiation - Fifteen to twenty (15-20) patients will receive a DR standard of care chest exam using the DRX Plus detector, and a DT exam. Each DT patient exam includes a scout image (chest PA) and a DT scan using the investigational DT SW. The DT scan is used by the DT console software to generate tomographic images.

Radiation: Radiation

Predicate & Invest. DT - Phantom Images

EXPERIMENTAL

Radiation - Eleven (11) phantoms of various anatomy will be imaged with linear tomography (LT) as predicate and DT for investigational.

Radiation: Radiation

Interventions

RadiationRADIATION

Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure. Phantoms were imaged using linear tomography and digital tomography.

Predicate & Invest. DE - Human SubjectsPredicate & Invest. DT - Human SubjectsPredicate & Invest. DT - Phantom Images

Eligibility Criteria

Age18 Years+
Sexall
Healthy VolunteersNo
Age GroupsAdult (18-64), Older Adult (65+)

You may qualify if:

  • Age 18 years and older. Subject has provided informed consent. Study participant is scheduled to have a chest CT exam. Study participant is able to stand and be still during the exams.

You may not qualify if:

  • Human subject specific:
  • Not able or willing to provide Informed Consent, or consent is withdrawn. Study participant is a female of child bearing age.

Contact the study team to confirm eligibility.

Sponsors & Collaborators

MeSH Terms

Conditions

Solitary Pulmonary Nodule

Interventions

Radiation

Condition Hierarchy (Ancestors)

Lung DiseasesRespiratory Tract Diseases

Intervention Hierarchy (Ancestors)

Physical Phenomena

Results Point of Contact

Title
Susan Pate, RN, MS, RAC, Senior Clinical Manager
Organization
Carestream Health, Inc.

Study Officials

  • Narinder Paul, MD

    Toronto General Hospital

    PRINCIPAL INVESTIGATOR

Publication Agreements

PI is Sponsor Employee
No
Restrictive Agreement
No

Study Design

Study Type
interventional
Phase
not applicable
Allocation
RANDOMIZED
Masking
DOUBLE
Who Masked
PARTICIPANT, OUTCOMES ASSESSOR
Purpose
DIAGNOSTIC
Intervention Model
PARALLEL
Sponsor Type
INDUSTRY
Responsible Party
SPONSOR

Study Record Dates

First Submitted

August 20, 2015

First Posted

August 24, 2015

Study Start

November 1, 2015

Primary Completion

April 1, 2016

Study Completion

April 1, 2016

Last Updated

March 9, 2017

Results First Posted

November 2, 2016

Record last verified: 2015-08

Data Sharing

IPD Sharing
Will not share