EMG-Based Hand-Wrist Control: Study B Mirrored
PSICON-B
3 other identifiers
interventional
20
1 country
2
Brief Summary
Study assessing four-channel prosthesis controller, that compares contralateral (mirrored) EMG-force training to ipsilateral EMG-target training with both limb-absent and able-bodied subjects
Trial Health
Trial Health Score
Automated assessment based on enrollment pace, timeline, and geographic reach
participants targeted
Target at P25-P50 for early_phase_1
Started Jun 2017
Typical duration for early_phase_1
2 active sites
Health score is calculated from publicly available data and should be used for screening purposes only.
Trial Relationships
Click on a node to explore related trials.
Study Timeline
Key milestones and dates
Study Start
First participant enrolled
June 29, 2017
CompletedPrimary Completion
Last participant's last visit for primary outcome
August 24, 2020
CompletedStudy Completion
Last participant's last visit for all outcomes
August 24, 2020
CompletedFirst Submitted
Initial submission to the registry
December 28, 2020
CompletedFirst Posted
Study publicly available on registry
January 5, 2021
CompletedJanuary 5, 2021
December 1, 2020
3.2 years
December 28, 2020
December 30, 2020
Conditions
Keywords
Outcome Measures
Primary Outcomes (1)
EMG-Force Models
Electromyogram waveform measurements (signal amplitude, frequency) will be related to force measurements (magnitude and moment) as related to the target tracking tasks. These signal-force data will be use to generate the models. The specific outcome for each subject will be the RMS error measured in the tracking tasks as the difference between the generated control signal from the user versus the prompted target signal, as measured by an EMG array.
1 day visit
Study Arms (2)
Study B: Limb-Absent Subject
EXPERIMENTALTen unilateral transradial limb-absent subjects will each participate in one, half-day experimental session. Subjects will be seated and prepared in test apparatus seat (16 electrodes on the affected side, hand-wrist on the able side secured to load cells). Subjects will complete the 1-DoF and 2-DoF "dynamic" (force-varying) contractions (40-s duration, 0.75 Hz bandlimited, uniform random target). With 2 DoF contraction trials, hand Opn-Cls will always be one of the dimensions. The subject will then be released from the cuff and their able side not further involved in the experiment. The force feedback triangle cursor on the computer screen will be deleted such that only the target remains. Subjects will then repeat 1-DoF and 2-DoF trials in which the affected side attempts to produce hand-wrist effort that mimic movement of the target (with no feedback provided).
Study B: Able-bodied Subjects
EXPERIMENTALTen able-bodied subjects, the electrodes will be mounted on the dominant arm for the EMG-force and EMG-target tracking trials. In addition, the non-dominant arm will also be constrained and measured in a second load cell. This load cell will not be used for feedback during the experiment, but will compare (RMS error, off-line) the dominant vs. non-dominant forces. For EMG-target tracking, the dominant hand will remain in the wrist cuff (to prevent flailing during contractions), with the screen feedback disabled. These subjects will repeat the trials with the electrodes moved to the non-dominant side. EMG-force tracking will be repeated using mirrored contractions. The three training methods (EMG-force ipsilateral, EMG-force contralateral mirrored, EMG-target on the dominant side) will be contrasted to help understand the source of errors when training with limb-absent subjects.
Interventions
Test apparatus acquires 16 sEMG channels, measure four DoFs of force/moment at the hand-wrist \[hand open-close (Opn-Cls); wrist extension-flexion (Ext-Flx), radial-ulnar deviation (Rad-Uln) and pronation-supination (Pro-Sup)\]. sEMG system used 16 custom encased bipolar electrodes. A commercial LTI EMG amplifier (BE328) conditioned each EMG signal before A/D conversion on a PC. A thermoplastic hand cuff secured the wrist to a six-axis load cell (AMTI, model MC3A-100) for force/moment measurement. A separate hand-grasp one-axis load cell (Omega Engineering Inc., model LCR-150) was Velcro-secured between the fingers and the thumb, to measure power grip forces. The PC acquired and stored the sEMG and load cell data, and commanded a triangular screen target. The subject controlled a second triangular cursor. Left-right movement of the cursor was controlled by wrist Ext-Flx, up-down movement by Rad-Uln deviation, cursor rotation by Pro-Sup and cursor size by hand Opn-Cls forces.
Eligibility Criteria
You may qualify if:
- Limb-absent subjects: unilateral trans-radial limb absence or amputation
- Be capable of completing the requested contractions on the affected dominant side
You may not qualify if:
- Past injuries to the upper limbs that would limit their ability to complete the requested contractions
- Scars that would impede the use of surface electrodes
Contact the study team to confirm eligibility.
Sponsors & Collaborators
Study Sites (2)
Liberating Technologies, Inc
Holliston, Massachusetts, 01746, United States
Worcester Polytechnic Institute
Worcester, Massachusetts, 01609, United States
MeSH Terms
Conditions
Study Design
- Study Type
- interventional
- Phase
- early phase 1
- Allocation
- NON RANDOMIZED
- Masking
- NONE
- Purpose
- BASIC SCIENCE
- Intervention Model
- SEQUENTIAL
- Sponsor Type
- INDUSTRY
- Responsible Party
- SPONSOR
Study Record Dates
First Submitted
December 28, 2020
First Posted
January 5, 2021
Study Start
June 29, 2017
Primary Completion
August 24, 2020
Study Completion
August 24, 2020
Last Updated
January 5, 2021
Record last verified: 2020-12