Improving Safety of Lead Extraction Procedures by Traction Force Sensing
1 other identifier
interventional
22
1 country
1
Brief Summary
Death and vascular tears occur in up to 1% of the patients undergoing lead extraction procedures. These complications are due to abnormal force vectors during the procedure. The investigators postulate that use of bidirectional traction (superior and inferior directions) and feedback from a digital force gauge is associated with minimal lead displacement compared to conventional lead extraction. Lead displacement is an indirect risk factor for potentially fatal lead extraction complications
Trial Health
Trial Health Score
Automated assessment based on enrollment pace, timeline, and geographic reach
participants targeted
Target at below P25 for not_applicable
Started Dec 2018
Typical duration for not_applicable
1 active site
Health score is calculated from publicly available data and should be used for screening purposes only.
Trial Relationships
Click on a node to explore related trials.
Study Timeline
Key milestones and dates
Study Start
First participant enrolled
December 20, 2018
CompletedFirst Submitted
Initial submission to the registry
January 9, 2019
CompletedFirst Posted
Study publicly available on registry
February 15, 2019
CompletedPrimary Completion
Last participant's last visit for primary outcome
December 10, 2020
CompletedStudy Completion
Last participant's last visit for all outcomes
December 10, 2020
CompletedJune 23, 2022
June 1, 2022
2 years
January 9, 2019
June 21, 2022
Conditions
Keywords
Outcome Measures
Primary Outcomes (1)
Fluoroscopic lead displacement
The primary objective of the study is to evaluate the effect of traction force as measured by a digital force gauge on fluoroscopic lead displacement during lead extraction procedures. The investigators hypothesize that the lead displacement on fluoroscopy will be less than 1 cm when the traction force is limited to less than 8 lb.
Intra-procedure measurements
Study Arms (2)
Intervention
ACTIVE COMPARATORUse of bidirectional rail (superior and inferior approaches) and force sensing using a force gauge to optimize Force application to less than 8 lbs during the extraction procedure.
Control
NO INTERVENTIONConventional lead extraction procedures using a superior approach is performed by experienced operators. Although force information is available the operators are blinded to the information. Inferior rail is left to the discretion of the operator.
Interventions
The investigators postulate that use of bidirectional traction (superior and inferior directions) and feedback from a digital force gauge is associated with minimal lead displacement compared to conventional lead extraction.
Eligibility Criteria
You may qualify if:
- Patients with CIED infection requiring extraction
You may not qualify if:
- Age \<18 years
- Noninfectious reasons for extraction
- Patients with cognitive impairment, who are unable to consent.
Contact the study team to confirm eligibility.
Sponsors & Collaborators
- Mayo Cliniclead
Study Sites (1)
Mayo Clinic in Rochester
Rochester, Minnesota, 55905, United States
Related Links
MeSH Terms
Conditions
Condition Hierarchy (Ancestors)
Study Officials
- PRINCIPAL INVESTIGATOR
Siva K Mulpuru, MD
Mayo Clinic
Study Design
- Study Type
- interventional
- Phase
- not applicable
- Allocation
- RANDOMIZED
- Masking
- SINGLE
- Who Masked
- OUTCOMES ASSESSOR
- Masking Details
- The study is randomized such that the same operator can be doing the extraction with force gauge with or without the force information.
- Purpose
- OTHER
- Intervention Model
- PARALLEL
- Sponsor Type
- OTHER
- Responsible Party
- PRINCIPAL INVESTIGATOR
- PI Title
- Principal Investigator
Study Record Dates
First Submitted
January 9, 2019
First Posted
February 15, 2019
Study Start
December 20, 2018
Primary Completion
December 10, 2020
Study Completion
December 10, 2020
Last Updated
June 23, 2022
Record last verified: 2022-06