Intramuscular Heating Rates of a Chattanooga Intelect XT Therapeutic Ultrasound
1 other identifier
interventional
25
1 country
1
Brief Summary
The purpose of this study is to determine the rate of tissue temperature increase in the triceps surae during an ultrasound treatment using the Chattanooga Intelect Legend XT. The following parameters will be used: 3 MHz frequency, 1.0 W/cm2 intensity, 15 minute treatment time. Thermocouples will be inserted at 1.0, 1.75, and 2.5 cm depths to determine the heating rates. Most modality textbook recommendations for therapeutic ultrasound are based on studies performed with the Omnisound ultrasound machine. Previous research has shown variations in power output across manufacturers, producing differences in tissue temperature increases. No information has been published regarding the heating rates of the Chattanooga Intelect Legend XT machine for ultrasound treatments. By determining the heating rates for this machine, clinicians will be better able to utilize the Chattanooga Intelect Legend XT to help reach therapeutic goals.
Trial Health
Trial Health Score
Automated assessment based on enrollment pace, timeline, and geographic reach
participants targeted
Target at below P25 for not_applicable
Started Feb 2019
1 active site
Health score is calculated from publicly available data and should be used for screening purposes only.
Trial Relationships
Click on a node to explore related trials.
Study Timeline
Key milestones and dates
First Submitted
Initial submission to the registry
January 23, 2019
CompletedFirst Posted
Study publicly available on registry
January 25, 2019
CompletedStudy Start
First participant enrolled
February 5, 2019
CompletedPrimary Completion
Last participant's last visit for primary outcome
April 15, 2020
CompletedStudy Completion
Last participant's last visit for all outcomes
May 11, 2020
CompletedResults Posted
Study results publicly available
May 27, 2021
CompletedMay 27, 2021
May 1, 2021
1.2 years
January 23, 2019
March 5, 2020
May 4, 2021
Conditions
Keywords
Outcome Measures
Primary Outcomes (3)
Heating Rate at 1 cm Depth
A thermocouple was inserted at the 1 cm depth to determine the tissue temperatures.
Baseline to end of treatment (up to 15 minutes or until temperatures increased 4 degrees, which occurred at 6 minutes)
Heating Rate at 1.75 cm Depth
A thermocouple was inserted at the 1.75 cm depth to determine the tissue temperatures.
Baseline to end of treatment (up to 15 minutes or until temperatures increased 4 degrees, which occurred at 8 minutes)
Heating Rate at 2.5 cm Depth
A thermocouple was inserted at the 2.5 cm depth to determine the tissue temperatures.
Baseline to end of treatment (up to 15 minutes or until temperatures increased 4 degrees, which did not occur in the 15 minute treatment)
Study Arms (1)
Therapeutic ultrasound treatment
OTHERAll subjects received a therapeutic ultrasound treatment with the Chattanooga Intelect Legend XT machine with the following parameters: 3 MHz, 1.0 W/cm2, 15 minute treatment time.
Interventions
A therapeutic ultrasound treatment was applied to the left medial gastrocnemius muscle with the following parameters: 3 MHz, 1.0 W/cm2 for 15 minutes.
Eligibility Criteria
You may qualify if:
- males and females between the ages of 18 - 35 years old
- able to speak English
- have less than 1 cm of adipose over the left medial gastrocnemius muscle.
You may not qualify if:
- any current lower extremity injury
- any lower extremity injuries within the last 6 months
- open wounds
- skin infections
- impaired circulation
- peripheral vascular disease
- deep vein thrombosis
- advanced arthritis
- poor thermal regulation
- anesthetic areas
- neoplasms/tumors
- thrombophlebitis
- closed infections
- any allergies or sensitivities to betadine
- any abnormalities observed in the medial gastrocnemius muscle determined by diagnostic ultrasound
- +1 more criteria
Contact the study team to confirm eligibility.
Sponsors & Collaborators
Study Sites (1)
North Dakota State University
Fargo, North Dakota, 58108-6050, United States
Related Publications (3)
Gange KN, Kjellerson MC, Berdan CJ. The Dynatron Solaris(R) Ultrasound Machine: Slower Heating Than Textbook Recommendations at 3 MHz, 1.0 W/cm2. J Sport Rehabil. 2018 Jan 1;27(1):22-29. doi: 10.1123/jsr.2016-0173. Epub 2018 Jan 18.
PMID: 27992304BACKGROUNDMerrick MA, Bernard KD, Devor ST, Williams MJ. Identical 3-MHz ultrasound treatments with different devices produce different intramuscular temperatures. J Orthop Sports Phys Ther. 2003 Jul;33(7):379-85. doi: 10.2519/jospt.2003.33.7.379.
PMID: 12918863BACKGROUNDDraper DO, Castel JC, Castel D. Rate of temperature increase in human muscle during 1 MHz and 3 MHz continuous ultrasound. J Orthop Sports Phys Ther. 1995 Oct;22(4):142-50. doi: 10.2519/jospt.1995.22.4.142.
PMID: 8535471BACKGROUND
MeSH Terms
Interventions
Intervention Hierarchy (Ancestors)
Results Point of Contact
- Title
- Kara Gange, Associate Professor
- Organization
- North Dakota State University
Study Officials
- STUDY DIRECTOR
Kristy Shirley
NDSU IRB
Publication Agreements
- PI is Sponsor Employee
- Yes
Study Design
- Study Type
- interventional
- Phase
- not applicable
- Allocation
- NA
- Masking
- NONE
- Purpose
- BASIC SCIENCE
- Intervention Model
- SINGLE GROUP
- Sponsor Type
- OTHER
- Responsible Party
- PRINCIPAL INVESTIGATOR
- PI Title
- Associate Professor
Study Record Dates
First Submitted
January 23, 2019
First Posted
January 25, 2019
Study Start
February 5, 2019
Primary Completion
April 15, 2020
Study Completion
May 11, 2020
Last Updated
May 27, 2021
Results First Posted
May 27, 2021
Record last verified: 2021-05
Data Sharing
- IPD Sharing
- Will not share
Once the data has been analyzed to determine the heating rates, a manuscript including the heating rates will be submitted for publication.