Clinical System Accuracy Evaluation With OneTouch® Verio Flex Blood Glucose Monitoring System (BGMS)
1 other identifier
interventional
109
1 country
1
Brief Summary
The objective of this study is the evaluation of system accuracy based on EN ISO 15197:2015 (ISO 15197:2013), clause 6.3 in which accuracy requirements and applicable test procedures for blood glucose monitoring systems intended for self-monitoring of blood glucose by patients are stipulated. ISO 15197:20132 was harmonized with the regulations of the European Union as EN ISO 15197:2015. This harmonization had no impact on the requirements and procedu¬res in ISO 15197:2013; changes were made to the foreword and an informative annex. In this study, system accuracy evaluation will be performed for OneTouch® Verio Flex (LifeScan Europe) on behalf of LifeScan Scotland Limited (customer) with three different reagent system lots.
Trial Health
Trial Health Score
Automated assessment based on enrollment pace, timeline, and geographic reach
participants targeted
Target at P50-P75 for not_applicable diabetes-mellitus
Started Mar 2017
Shorter than P25 for not_applicable diabetes-mellitus
1 active site
Health score is calculated from publicly available data and should be used for screening purposes only.
Trial Relationships
Click on a node to explore related trials.
Study Timeline
Key milestones and dates
First Submitted
Initial submission to the registry
March 15, 2017
CompletedStudy Start
First participant enrolled
March 16, 2017
CompletedFirst Posted
Study publicly available on registry
March 21, 2017
CompletedPrimary Completion
Last participant's last visit for primary outcome
April 3, 2017
CompletedStudy Completion
Last participant's last visit for all outcomes
June 1, 2017
CompletedSeptember 26, 2017
September 1, 2017
18 days
March 15, 2017
September 25, 2017
Conditions
Outcome Measures
Primary Outcomes (1)
System accuracy criteria
Based on EN ISO 15197:2015 (ISO 15197:2013) (E), the BGMS shall meet both of the following criteria: Criterion A: 95 % of the measured glucose values shall fall within either ± 15 mg/dl (0.83 mmol/l) of the average comparison measurement result at glucose concentrations \< 100 mg/dl (5.55 mmol/l) or within ± 15 % at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l). Each lot shall pass acceptability criterion A. Criterion B: 99 % of individual glucose measured values shall fall within zones A and B of the Consensus Error Grid (CEG) for type 1 diabetes.
For each subject, the experimental phase has an expected duration of up to 6 hours
Study Arms (1)
Blood glucose measurement
EXPERIMENTALInterventions
Measurement procedure * Subjects will be asked to wash and dry their hands before the measurements. * The measurement procedure for each reagent system lot of a BGMS will be as follows: * Study personnel will collect a capillary blood sample from the subject's fingertip by skin puncture. * Duplicate measurements on the sample will be consecutively performed with 2 test meters using reagent system units from the same vial / package. * Capillary blood for measurements with all 3 test strip lots and for both comparison measurements will be taken from the same finger puncture. * The sample will be applied to the reagent system units following specifications in the manufacturer's labelling.
Eligibility Criteria
You may qualify if:
- Male or female, with type 1 diabetes, type 2 diabetes (or subjects without diabetes)
- Signed informed consent form
- Minimum age of 18 years
- Subjects are legally competent and capable to understand character, meaning and consequences of the study.
- If blood glucose values \< 80 mg/dl or \> 300 mg/dl shall be measured after short term alteration in insulin therapy:
- Male or female with type 1 diabetes and intensified insulin therapy or insulin pump therapy.
- Signature of subjects to document consent with these procedures on informed consent form.
You may not qualify if:
- Pregnancy or lactation period
- Severe acute disease (at the study physician's discretion)
- Severe chronic disease with potential risk during the test procedures (at the study physician's discretion)
- Current constitution that compromises the subject's capability to participate in the study (at the study physician's discretion)
- Being unable to give informed consent
- \< 18 years
- Legally incompetent
- Being committed to an institution (e.g. psychiatric clinic)
- Language barriers potentially compromising an adequate compliance with study procedures
- Dependent on investigator or sponsor
- If blood glucose values \< 80 mg/dl shall be measured after short term alteration in insulin therapy (concentration category 2), subjects with type 1 diabetes, suffering from:
- Coronary heart disease
- Condition after myocardial infarction
- Condition after cerebral events
- Peripheral arterial occlusive disease
- +1 more criteria
Contact the study team to confirm eligibility.
Sponsors & Collaborators
Study Sites (1)
Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm
Ulm, Baden-Wurttemberg, 89081, Germany
MeSH Terms
Conditions
Interventions
Condition Hierarchy (Ancestors)
Intervention Hierarchy (Ancestors)
Study Design
- Study Type
- interventional
- Phase
- not applicable
- Allocation
- NA
- Masking
- NONE
- Purpose
- OTHER
- Intervention Model
- SINGLE GROUP
- Sponsor Type
- OTHER
- Responsible Party
- SPONSOR
Study Record Dates
First Submitted
March 15, 2017
First Posted
March 21, 2017
Study Start
March 16, 2017
Primary Completion
April 3, 2017
Study Completion
June 1, 2017
Last Updated
September 26, 2017
Record last verified: 2017-09