NCT03054649

Brief Summary

The purpose of this study was to demonstrate the safety and effectiveness of the HARMONI® Modular Intraocular Lens System with a toric optic in subjects with pre-existing corneal astigmatism in need of cataract surgery.

Trial Health

90
On Track

Trial Health Score

Automated assessment based on enrollment pace, timeline, and geographic reach

Enrollment
44

participants targeted

Target at P25-P50 for not_applicable

Timeline
Completed

Started Jun 2016

Geographic Reach
2 countries

2 active sites

Status
completed

Health score is calculated from publicly available data and should be used for screening purposes only.

Trial Relationships

Click on a node to explore related trials.

Study Timeline

Key milestones and dates

Study Start

First participant enrolled

June 28, 2016

Completed
8 months until next milestone

First Submitted

Initial submission to the registry

February 12, 2017

Completed
3 days until next milestone

First Posted

Study publicly available on registry

February 15, 2017

Completed
5 months until next milestone

Primary Completion

Last participant's last visit for primary outcome

July 18, 2017

Completed
Same day until next milestone

Study Completion

Last participant's last visit for all outcomes

July 18, 2017

Completed
2.9 years until next milestone

Results Posted

Study results publicly available

June 22, 2020

Completed
Last Updated

June 22, 2020

Status Verified

June 1, 2020

Enrollment Period

1.1 years

First QC Date

February 12, 2017

Results QC Date

June 2, 2020

Last Update Submit

June 2, 2020

Conditions

Keywords

cataractintraoperative exchangeintraocular lensastigmatismtoric

Outcome Measures

Primary Outcomes (17)

  • Mean Manifest Refraction Cylinder (MRCYL) for Eyes Implanted With HMTIOL

    A manifest refraction (manual vision test) was conducted using letter charts and a phoropter. Manifest refraction cylinder is the amount of added correction needed to compensate for any astigmatism that may be present. MRCYL was measured in diopters and reported as a negative number, with a less negative number indicating a lesser amount of added correction. No formal statistical hypothesis testing was planned.

    Month 1 postoperative, Month 3 postoperative

  • Mean Manifest Refraction Cylinder (MRCYL) Prediction Error (PE) for Eyes Implanted With HMTIOL

    A manifest refraction (manual vision test) was conducted using letter charts and a phoropter. Manifest refraction cylinder is the amount of power needed to correct any astigmatism that may be present. MRCYL was measured in diopters and reported as a negative number. MRCYL PE was calculated as postoperative MRCYL adjusted to 6 meters minus MRCYL target residual refractive error (TRRE), with a lower absolute value indicating a more accurate cylinder power calculation. No formal statistical hypothesis testing was planned.

    Month 1 postoperative, Month 3 postoperative

  • Mean Reduction in Cylinder Power for Eyes Implanted With HMTIOL

    A manifest refraction (manual vision test) was conducted using letter charts and a phoropter. Manifest refraction cylinder is the amount of power needed to correct any astigmatism that may be present. MRCYL was measured in diopters and reported as a negative number. Cylinder power reduction was calculated as the absolute preoperative magnitude of corneal cylinder (K) minus the absolute postoperative magnitude of MRCYL at the corneal plane. No formal statistical hypothesis testing was planned.

    Baseline (Day 0 preoperative), Month 3 postoperative

  • Mean Percent Reduction in Cylinder Power for Eyes Implanted With HMTIOL

    A manifest refraction (manual vision test) was conducted using letter charts and a phoropter. Manifest refraction cylinder is the amount of power needed to correct any astigmatism that may be present. MRCYL was measured in diopters and reported as a negative number. Cylinder power reduction was calculated as the absolute preoperative magnitude of corneal cylinder (K) minus the absolute postoperative magnitude of MRCYL at the corneal plane. Therefore, the cylindrical power percent reduction was the cylinder power reduction divided by the absolute value of preoperative magnitude of corneal cylinder (K) × 100. A higher number (greater percent reduction) indicates better astigmatism correction by the toric IOL. No formal statistical hypothesis testing was planned.

    Baseline (Day 0 preoperative), Month 3 postoperative

  • Mean Manifest Refraction Spherical Equivalent (MRSE) Prediction Error (PE)

    A manifest refraction (manual vision test) was conducted using letter charts and a phoropter. The manifest refraction spherical equivalent (MRSE) was calculated as follows: sphere + 1/2 cylinder, and measured in diopters (D). MRSE PE was calculated as postoperative MRSE adjusted to 6 meters minus MRSE target residual refractive error (TRRE). The lower the absolute number, the more accurate the IOL power calculation is. No formal statistical hypothesis testing was planned.

    Month 3 postoperative

  • Mean Absolute Rotation of IOL Meridian

    IOL orientation was measured with slit-lamp photography. IOL rotation was defined as the difference between axis of IOL orientation on the day of surgery and the subsequent postoperative visit. IOL rotation was measured in degrees, with a lower number indicating less rotation. No formal statistical hypothesis testing was planned.

    Baseline (Day 0 operative), Day 1 postoperative, Week 1 postoperative, Month 1 postoperative, Month 3 postoperative

  • Number of Eyes With Absolute Rotation of IOL Meridian by Visit - Cohort 1

    IOL orientation was measured with slit-lamp photography. IOL rotation was defined as the difference between axis of IOL orientation on the day of surgery and the subsequent postoperative visit. IOL rotation was measured in degrees and is reported categorically. No formal statistical hypothesis testing was planned.

    Baseline (Day 0 operative), Day 1 postoperative, Week 1 postoperative, Month 1 postoperative, Month 3 postoperative

  • Number of Eyes With Absolute Rotation of IOL Meridian by Visit - Cohort 2

    IOL orientation was measured with slit-lamp photography. IOL rotation was defined as the difference between axis of IOL orientation on the day of surgery and the subsequent postoperative visit. IOL rotation was measured in degrees and is reported categorically. No formal statistical hypothesis testing was planned.

    Baseline (Day 0 operative), Day 1 postoperative, Week 1 postoperative, Month 1 postoperative, Month 3 postoperative

  • Percentage of Eyes With Best Corrected Distance Visual Acuity (BCDVA) by Visit - Cohort 1

    Visual Acuity of the eye was tested while reading charts at 20-foot equivalent distance from the participant with the correction obtained from manifest refraction testing. BCDVA is reported categorically using the Snellen fraction, which compares the participant's result to the result expected from a 'normal' visual system. The numerator represents the distance between the participant and the chart, and the denominator represents the distance at which a person with 'normal' vision would be able to discern the same letter size. 20/20 is considered to be 'normal' vision, whereas visual acuity of 20/40 means the participant is able to read a certain size letter 20 feet away that a person with 'normal' vision would be able to read from 40 feet away. A larger denominator, therefore, indicates a lower visual acuity. No formal statistical hypothesis testing was planned.

    Baseline (Day 0 operative), Week 1 postoperative, Month 1 postoperative, Month 3 postoperative

  • Percentage of Eyes With Best Corrected Distance Visual Acuity (BCDVA) by Visit - Cohort 2

    Visual Acuity of the eye was tested while reading charts at 20-foot equivalent distance from the participant with the correction obtained from manifest refraction testing. UCDVA is reported categorically using the Snellen fraction, which compares the participant's result to the result expected from a 'normal' visual system. The numerator represents the distance between the participant and the chart, and the denominator represents the distance at which a person with 'normal' vision would be able to discern the same letter size. 20/20 is considered to be 'normal' vision, whereas visual acuity of 20/40 means the participant is able to read a certain size letter 20 feet away that a person with 'normal' vision would be able to read from 40 feet away. A larger denominator, therefore, indicates a lower visual acuity. No formal statistical hypothesis testing was planned.

    Baseline (Day 0 operative), Week 1 postoperative, Month 1 postoperative, Month 3 postoperative

  • Percentage of Eyes With Uncorrected Distance Visual Acuity (UCDVA) by Visit - Cohort 1

    Visual Acuity of the eye was tested while reading charts at 20-foot equivalent distance from the participant with an optical infinity adjustment of +0.25 diopter (D). The Snellen fraction compares the participant's result to the result expected from a 'normal' visual system. The numerator represents the distance between the participant and the chart, and the denominator represents the distance at which a person with 'normal' vision would be able to discern the same letter size. 20/20 is considered to be 'normal' vision, whereas visual acuity of 20/40 means the participant is able to read a certain size letter 20 feet away that a person with 'normal' vision would be able to read from 40 feet away. A larger denominator, therefore, indicates a lower visual acuity. No formal statistical hypothesis testing was planned.

    Baseline (Day 0 operative), Day 1 postoperative, Week 1 postoperative, Month 1 postoperative, Month 3 postoperative

  • Percentage of Eyes With Uncorrected Distance Visual Acuity (UCDVA) by Visit - Cohort 2

    Visual Acuity of the eye was tested while reading charts at 20-foot equivalent distance from the participant with an optical infinity adjustment of +0.25 diopter (D). The Snellen fraction compares the participant's result to the result expected from a 'normal' visual system. The numerator represents the distance between the participant and the chart, and the denominator represents the distance at which a person with 'normal' vision would be able to discern the same letter size. 20/20 is considered to be 'normal' vision, whereas visual acuity of 20/40 means the participant is able to read a certain size letter 20 feet away that a person with 'normal' vision would be able to read from 40 feet away. A larger denominator, therefore, indicates a lower visual acuity. No formal statistical hypothesis testing was planned.

    Baseline (Day 0 operative), Day 1 postoperative, Week 1 postoperative, Month 1 postoperative, Month 3 postoperative

  • Number of Eyes With Preservation of BCDVA at Month 1 and Month 3

    Preservation of BCDVA was defined in terms of lines lost when reading the ETDRS chart at each visit compared to the prior best-achieved BCDVA. Loss of \< 1 line included subjects with less than a 1 line loss, no change or improvement in BCDVA. No formal statistical hypothesis testing was planned.

    Week 1 postoperative, Month 1 postoperative, Month 3 postoperative

  • Number of Ocular Adverse Events Through Month 3

    Ocular adverse events were identified as outlined in ISO 11979-2014 Annex B. No formal statistical hypothesis testing was planned.

    Up to Month 3 postoperative

  • Number of Device Deficiencies Post Implantation

    A Device Deficiency (DD) was defined as a failure of the device to meet its performance specifications or expectations, or otherwise not perform as intended. This could include either a malfunction or damage to the device or any part thereof, regardless of the source of malfunction or damage, including user error, and regardless of the presence of injury (or lack thereof) to subject, user, or bystander. No formal statistical hypothesis testing was planned.

    Up to Month 3 postoperative

  • Number of Secondary Surgical Interventions, Other Than Intraroperative Optic Exchange and Rotational Adjustment of the Toric Optic in Cohort 2

    A secondary surgical intervention (SSI) was defined as a surgical procedure that occurred after primary implantation and was conducted for the purpose of resolving residual refractive error (RRE) and optimizing visual outcomes. No formal statistical hypothesis testing was planned.

    Up to Month 3 postoperative

  • Percent Change in Endothelial Cell Count (ECC) at Month 3

    High quality images of the corneal endothelium (the single layer of cells on the inner surface of the cornea) were taken at the investigative site and sent to an independent central reading center for evaluation. A negative percent change value may indicate a decrease in the ability of the endothelium to maintain its primary function. No formal statistical hypothesis testing was planned.

    Baseline (preoperative), Month 3 postoperative

Study Arms (2)

Cohort 1

EXPERIMENTAL

Study eye: Primary implantation of HMIOL system with toric optic and no intraoperative optic exchange

Device: HARMONI® Modular Intraocular Lens System

Cohort 2

EXPERIMENTAL

Fellow eye: Primary implantation of HMIOL system with monofocal optic, followed by intraoperative optic exchange (toric or non-toric optic)

Device: HARMONI® Modular Intraocular Lens SystemProcedure: Intraoperative optic exchange

Interventions

Two-component system consisting of a base and a separate optic that allows for an intraoperative exchange of the optic only without direct manipulation of the capsular bag

Also known as: HMIOL
Cohort 1Cohort 2

Removal of one optic and replacement with another during cataract surgery for the purpose of improving refractive outcomes

Also known as: IOE
Cohort 2

Eligibility Criteria

Age22 Years+
Sexall
Healthy VolunteersNo
Age GroupsAdult (18-64), Older Adult (65+)

You may qualify if:

  • Willing and able to return for scheduled treatment and follow-up examinations for up to 6 month study duration.
  • Planned removal of visually significant bilateral cataracts.
  • Pre-existing corneal astigmatism in at least 1 eye.
  • Willing to discontinue contact lens wear for the duration of the study.
  • Best Corrected Visual Acuity (BCVA) projected to be 0.2 logarithm Minimum Angle of Resolution (logMAR) or lower.
  • Visual symptoms related to cataracts.

You may not qualify if:

  • History of any intraocular or corneal surgery in either eye (including refractive).
  • Pregnant or lactating.
  • History of any clinically significant retinal pathology or ocular diagnosis in either eye that could alter or limit final postoperative visual prognosis.
  • History of any ocular condition which could affect the stability of the intraocular lens (IOL) in either eye.
  • Uncontrolled glaucoma in either eye (per Investigator judgement).
  • Severe dry eye that, in the opinion of the investigator, would impair the ability to obtain reliable study measurements

Contact the study team to confirm eligibility.

Sponsors & Collaborators

Study Sites (2)

ClarVista Investigative Site

Tijuana, Estado de Baja California, 22320, Mexico

Location

ClarVista Investigative Site

Makati, Manila, 1200, Philippines

Location

MeSH Terms

Conditions

CataractAstigmatismAphakia

Condition Hierarchy (Ancestors)

Lens DiseasesEye DiseasesRefractive Errors

Results Point of Contact

Title
Sr. Clinical Trial Lead, CDMA Surgical
Organization
Alcon Research, LLC

Study Officials

  • Sr. Clinical Trial Lead, CDMA Surgical

    Alcon Research, LLC

    STUDY DIRECTOR

Publication Agreements

PI is Sponsor Employee
No
Restriction Type
OTHER
Restrictive Agreement
Yes

Study Design

Study Type
interventional
Phase
not applicable
Allocation
NON RANDOMIZED
Masking
NONE
Purpose
TREATMENT
Intervention Model
SINGLE GROUP
Sponsor Type
INDUSTRY
Responsible Party
SPONSOR

Study Record Dates

First Submitted

February 12, 2017

First Posted

February 15, 2017

Study Start

June 28, 2016

Primary Completion

July 18, 2017

Study Completion

July 18, 2017

Last Updated

June 22, 2020

Results First Posted

June 22, 2020

Record last verified: 2020-06

Data Sharing

IPD Sharing
Will not share