Brief Summary
The purpose of this study is to compare difference of displacement between Polyvinyl Chloride(PVC) and Silicon double-lumen endobronchial tubes(DLTs) during changing position from supine to lateral in thoracic surgery.
Trial Health
Trial Health Score
Automated assessment based on enrollment pace, timeline, and geographic reach
participants targeted
Target at P50-P75 for not_applicable
Started Feb 2016
1 active site
Health score is calculated from publicly available data and should be used for screening purposes only.
Trial Relationships
Click on a node to explore related trials.
Study Timeline
Key milestones and dates
Study Start
First participant enrolled
February 1, 2016
CompletedFirst Submitted
Initial submission to the registry
February 5, 2016
CompletedFirst Posted
Study publicly available on registry
February 25, 2016
CompletedPrimary Completion
Last participant's last visit for primary outcome
April 1, 2017
CompletedStudy Completion
Last participant's last visit for all outcomes
May 1, 2017
CompletedResults Posted
Study results publicly available
December 23, 2019
CompletedAugust 10, 2021
August 1, 2021
1.2 years
February 5, 2016
May 25, 2018
August 9, 2021
Conditions
Keywords
Outcome Measures
Primary Outcomes (1)
the Incidence of Clinically Significant Displacement of DLTs During Change of Patient Position
The tracheal distance was defined as the distance between the distal tip of the tracheal lumen and tracheal carina whereas the bronchial distance was defined as that between the bronchial carina and distal tip of endobronchial lumen. The displacement of DLT was determined by changes in tracheal and bronchial distances, obtained by subtracting supine measurements from lateral measurements and subtracting measurements at start of surgery from measurements at the end of surgery, respectively. Clinically significant displacement was defined when the DLT was deviated by more than 10 mm from the initial correct position, regardless of the direction of displacement.
from supine to lateral decubitus position during surgery
Secondary Outcomes (1)
the Incidence of Critical Malposition of DLTs
from supine to lateral decubitus position
Study Arms (2)
PVC DLT
ACTIVE COMPARATORAfter the induction of general anesthesia, a left sided PVC DLT was placed by the anesthesiologist using direct laryngoscopy. Tube position was confirmed with fiberotic bronchoscope. The first set of measurements tracheal distance(from tracheal carina to proximal tip of DLT) and bronchial distance(from bronchial carina to distal tip of DLT) measured by bronchoscope is taken in supine position . The second set of measurements is taken in lateral position after position change. The displacement of PVC DLT is calculated from difference in tracheal distance between supine and lateral position.
silicon DLT
ACTIVE COMPARATORAfter the induction of general anesthesia, a left sided silicon DLT was placed by the anesthesiologist using direct laryngoscopy. Tube position was confirmed with fiberotic bronchoscope. The first set of measurements tracheal distance(from tracheal carina to proximal tip of DLT) and bronchial distance(from bronchial carina to distal tip of DLT) measured by bronchoscope is taken in supine position . The second set of measurements is taken in lateral position after position change. The displacement of silicon DLT is calculated from difference in tracheal distance between supine and lateral position.
Interventions
PVC DLT is type of double lumen tube that is composed of PVC
silicon DLT is type of double lumen tube that is composed of silicon
Eligibility Criteria
You may qualify if:
- Patient required differential lung ventilation in elective thoracic surgery American society of Anesthesiologist physical status(ASA PS) 1\~3
You may not qualify if:
- emergency surgery, difficult intubation, poor lung function to accomplish OLV during surgery
Contact the study team to confirm eligibility.
Sponsors & Collaborators
Study Sites (1)
Yeungnam University Hospital
Daegu, 42415, South Korea
Related Publications (3)
Benumof JL, Partridge BL, Salvatierra C, Keating J. Margin of safety in positioning modern double-lumen endotracheal tubes. Anesthesiology. 1987 Nov;67(5):729-38. doi: 10.1097/00000542-198711000-00018.
PMID: 3674473BACKGROUNDDesiderio DP, Burt M, Kolker AC, Fischer ME, Reinsel R, Wilson RS. The effects of endobronchial cuff inflation on double-lumen endobronchial tube movement after lateral decubitus positioning. J Cardiothorac Vasc Anesth. 1997 Aug;11(5):595-8. doi: 10.1016/s1053-0770(97)90011-2.
PMID: 9263092BACKGROUNDYoon TG, Chang HW, Ryu HG, Kwon TD, Bahk JH. Use of a neck brace minimizes double-lumen tube displacement during patient positioning. Can J Anaesth. 2005 Apr;52(4):413-7. doi: 10.1007/BF03016286.
PMID: 15814758BACKGROUND
Related Links
Limitations and Caveats
First, the position change and surgical manipulation may have a combined. Second, we did not monitor the pressure and volume of air injected into the bronchial cuff. Third, we did not investigate ideal position of DLT.
Results Point of Contact
- Title
- Sung Mee Jung, M.D
- Organization
- Department of Anesthesiology and Pain medicine, Yeungnam University School of Medicine, Daegu, Republic of Korea
Study Officials
- PRINCIPAL INVESTIGATOR
Sung Mee applejsm@gmail.com, MD
Yeungnam University College of Medicine
Publication Agreements
- PI is Sponsor Employee
- No
- Restrictive Agreement
- No
Study Design
- Study Type
- interventional
- Phase
- not applicable
- Allocation
- RANDOMIZED
- Masking
- SINGLE
- Who Masked
- PARTICIPANT
- Purpose
- PREVENTION
- Intervention Model
- PARALLEL
- Sponsor Type
- OTHER
- Responsible Party
- PRINCIPAL INVESTIGATOR
- PI Title
- Associate Professor
Study Record Dates
First Submitted
February 5, 2016
First Posted
February 25, 2016
Study Start
February 1, 2016
Primary Completion
April 1, 2017
Study Completion
May 1, 2017
Last Updated
August 10, 2021
Results First Posted
December 23, 2019
Record last verified: 2021-08
Data Sharing
- IPD Sharing
- Will not share